Typical Applications (Continued)
Single Power Supply Cold Junction Compensation
DS005698-11
*
Select R3 and R4 for thermocouple type
THERMO- R3 R4 SEEBECK
COUPLE COEFFICIENT
J 1.05K 385Ω 52.3 µV/˚C
T 856Ω 315Ω 42.8 µV/˚C
K 816Ω 300Ω 40.8 µV/˚C
S 128Ω 46.3Ω 6.4 µV/˚C
Adjustments:
1. Adjust R1 for the voltage across R3 equal to the Seebeck Coefficient times ambient temperature in degrees Kelvin.
2. Adjust R2 for voltage across R4 corresponding to thermocouple
J 14.32 mV
T 11.79 mV
K 11.17 mV
S 1.768 mV
LM135/LM235/LM335, LM135A/LM235A/LM335A
www.national.com9